Particle Analysis Applications in Materials Sciences
CMP Slurry Outlier Analysis with MFI
Chemical-Mechanical Planerization (CMP) requires precise control of the size distribution and morphology of particles contained in polishing slurries. A critical aspect affecting slurry performance is the presence of minute concentrations of large particles that can cause defects, such as micro-scratches, on the wafer during CMP.
These particles can originate either during the original manufacture or subsequently by aggregation during storage or use. The size, shape and degree of compaction of these large outlier particles will determine their scratching behavior. Micro-Flow Imaging (MF) uses digital imaging of a flowing sample stream for analyzing outlier particles in CMP slurries. MFI instruments provide highly sensitive detection, counting, and sizing of each individual particle combined with automatic image capture and shape analysis. This feature set results in instruments which can contribute to process yield enhancement both through more rigorous incoming and in-process inspection and, through the additional insight gained for process development and troubleshooting.
The MFI Advantage
When compared with alternative technologies for analyzing outlier populations in CMP slurries (principally light-blocking and light-scattering techniques as well as ultrasonic spectroscopy), MFI has a number of unique advantages including:
- Detection Sensitivity: Each particle is individually detected and measured using a sensitive, pixel-based imaging technique. This allows very low concentrations of outlier particles to be detected and measured. MFI is particularly well suited for the critical 0.75 to 10 µm range, accurately measuring outlier concentrations less than 0.001% by volume that are difficult to detect using existing techniques.
- Material and Shape Independence: Unlike many indirect particle size measurement technologies, measurements are insensitive to shape and composition making MFI well-suited for heterogeneous outlier populations commonly found in CMP slurries.
- Comprehensive Morphological Analysis: Images of outlier particles within the slurry population can be automatically stored and analyzed using the MFI morphological analysis module, providing measurement of equivalent circular diameter, Ferets diameter, area, perimeter, circularity, aspect ratio and transparency. These results can be conveniently presented in the form of histograms or scatterplots. This analysis provides further insight into the nature of the outlier particles as well as their propensity for damaging wafers.
- Speed and Convenience: MFI analyzers are easy to operate and provide full population analysis in less than 5 minutes.
- Direct Observation: During analysis, image frames are displayed providing immediate visual feedback on the nature of a particle population. Software assisted filtering can save particle images meeting user-defined characteristics.
MFI Sizing, Counting, Image Capture
The DPA4100 MFI instrument supports operation at two different magnification set points depending on the size range of interest (0.75µm to 100µm), or (2.25µm to 300µm). Images can be stored selectively based on detection of particles falling into user-specified size ranges, or sequentially as a function of time. The results below are an example of the DPA4100 analysis of two different cerium oxide slurry samples where one is known to be more problematic in terms of wafer damage. The system was found to be capable of measuring very small concentrations of particles larger than 5µm (0.001% by volume).